1300 µL clear reagent pipette tips for approximately 20–1000 µL transfers on compatible DYNEX DS2, DS2 Elite and DSX systems.
Puretest 1300 µL reagent pipette tips are designed as an independent compatible consumable for reagent transfer on DYNEX DS2, DS2 Elite and DSX automated ELISA systems. The clear large-capacity tip supports the instrument reagent-pipetting workflow while allowing direct visual inspection of the liquid column.
The 1300 µL designation describes the tip capacity; the instrument reagent-pipetting range is approximately 20–1000 µL. This distinction matters for method setup and prevents the incorrect assumption that the full physical tip capacity is the recommended dispense range. The 1300 µL reagent tip is separate from the blue 300 µL sample-tip family.
Important: These are independent compatible consumables, not original DYNEX consumables. Confirm the exact instrument, software method, pipetting role and physical configuration, then qualify the selected item before routine, regulated or diagnostic use.
The 1300 µL reagent tip uses a ribbed upper collar and long clear tapered shaft. The cone must seal consistently on the reagent-pipetting adapter while the shaft provides reach into reagent bottles, tubes or intermediate vessels. Sufficient internal capacity above the 1000 µL method maximum provides the geometry required by the instrument-specific format.
Reagent methods may involve larger volumes, repeated dispensing and liquids with different viscosities or surfactant content. Qualification should include full-volume aspiration, multi-dispense steps if used, residual-volume checks, reagent-vessel clearance and clean tip ejection. The exact carrier or supply tray must also be confirmed because instrument deck capacity is not the same as commercial packaging quantity.
| Parameter | Specification | Selection note |
|---|---|---|
| Nominal tip capacity | 1300 µL | Physical reagent-tip capacity. |
| Recommended working range | Approximately 20–1000 µL | Instrument reagent-pipetting guide; validate actual method volumes. |
| Instrument role | Reagent aspiration and dispensing | Not a 300 µL sample tip. |
| Tip color | Clear / natural | Visual inspection of the liquid column. |
| Tip material | Polypropylene | Confirm quality status and supplied packaging. |
| Carrier / packaging | Configuration-specific | Confirm carrier, position count and lid or tray before ordering. |
| Primary platforms | DYNEX DS2, DS2 Elite and DSX | Exact hardware and software method require qualification. |
| Item | Compatibility guidance | What to verify |
|---|---|---|
| DS2 / DS2 Elite | 1300 µL reagent-tip family; instrument reagent-tip capacity is 20 positions | Reagent-tip carrier, pickup path and method definition. |
| DSX | 1300 µL reagent-tip family; instrument reagent-tip capacity is 41 positions | Carrier fit, reagent-vessel access and software method. |
| Pipetting role | Approximately 20–1000 µL reagent transfer | Single or repeated dispense steps and representative reagents. |
| 300 µL sample tips | Separate blue sample-tip family | Do not load into the reagent-tip position or copy sample-tip settings. |
The two DYNEX-compatible tip families perform different functions and must not share copied specifications. The 300 µL format is the sample tip; the 1300 µL format is the reagent tip. Select the tip according to the programmed pipetting role, not only by visual shape or maximum capacity.
| Tip family | Instrument role | Initial working-volume guide |
|---|---|---|
| 300 µL sample tip | Sample, standard and control transfer | 10–300 µL |
| 1300 µL reagent tip | Reagent transfer and dispensing | 20–1000 µL |
For an accurate quotation, provide the exact DYNEX model, 1300 µL reagent-tip requirement, instrument carrier or tray details, required sterility/purity status, requested documents, package quantity and annual demand. Puretest supports evaluation samples, recurring bulk supply, distributor programs and OEM/private-label projects for confirmed configurations.
1300 µL is the nominal physical tip capacity; approximately 20–1000 µL is the reagent-pipetting range used for instrument selection and method qualification.
No. The 1300 µL clear tip is the reagent-tip family, while the 300 µL blue tip is used for sample pipetting.
Official system information lists 20 reagent-tip positions for DS2/DS2 Elite and 41 for DSX. Commercial packaging must be confirmed separately.
Potentially, when the instrument method supports it; validate reagent properties, total aspirated volume, residual volume and dispense consistency.
No. They are independent compatible tips and require exact-system and assay qualification.
Send the exact instrument model, required tip role, programmed volume range, carrier or tray details, required quality documents and estimated annual demand. We will match the closest configuration for evaluation.
Contact Puretest for a compatibility sample or quotation
DYNEX, DS2 and DSX are trademarks of their respective owner. Puretest is an independent compatible-consumables supplier and is not affiliated with or endorsed by DYNEX Technologies, Inc.
General Inquiries
Please leave your message here! We willsend detailed technical info and quotationto you!